This guide brings together CDSCO licensing requirements for 51 medical devices sold in India — from blood bags and nebulizers to infusion pumps and orthopaedic implants. For each device you will find its risk class under the Medical Devices Rules, 2017, the standards CDSCO expects, the documents specific to that device and the query that most often delays approval. For the full service, see our CDSCO registration services.
Quick answer: every medical device in India needs a CDSCO licence. Class A and B devices made in India are licensed by the State Licensing Authority (Form MD-3 → MD-5); Class C and D by the Central Licensing Authority (Form MD-7 → MD-9). Imported devices of any class need an import licence (Form MD-14 → MD-15) filed by an Indian Authorised Agent.
🧭 How CDSCO Classifies Medical Devices
The First Schedule of the Medical Devices Rules, 2017 places every device in one of four risk classes. The class decides the licensing authority, the form, the fee, whether a Notified Body audits the site before grant, and the timeline — so it is the first thing to settle.
🪪 Licensing Routes: Manufacture vs Import
- Manufacture, Class A/B: apply in Form MD-3 to the State Licensing Authority; licence granted in Form MD-5. A CDSCO-empanelled Notified Body audits the site (post-grant for Class A, pre-grant for Class B).
- Manufacture, Class C/D: apply in Form MD-7 to the Central Licensing Authority; licence granted in Form MD-9 after inspection.
- Import, any class: the overseas manufacturer appoints an Indian Authorised Agent holding a manufacturing or wholesale licence, who files Form MD-14; the import licence is granted in Form MD-15.
- Test licence: small quantities for testing, evaluation, demonstration or clinical investigation use Form MD-12 → MD-13.
💰 Fees, Timeline & Validity
Fees are set in the Second Schedule of the Medical Devices Rules and paid online, charged per site and per device. A Class B manufacturing licence, for example, costs Rs 5,000 per site plus Rs 500 per distinct device; import fees are paid in USD and rise with the class. A well-prepared manufacturing file typically clears in 3–6 months and an import file in 6–9 months. Licences are perpetual, subject to a retention fee every five years.
📑 Device Index
Class A: Examination Gloves, Surgical Masks
Class B: Autoclaves & Sterilizers, Blood Pressure Monitors, Condoms, Contact Lenses, Digital Thermometers, Disposable Syringes, ECG Machines, Enteral Feeding Tubes, Hearing Aids, Hospital Beds, Insulin Pens, IV Cannula, Nebulizers, Oxygen Concentrators, Pregnancy Test Kits, Surgical Gloves, Ultrasound Machines, Urinary Catheters, Wound Dressings
Class C: Anaesthesia Machines, Blood Bags, Blood Gas Analysers, Cardiac Catheters, COVID-19 Antigen Test Kits, CT Scanners, Defibrillators, Dengue & Malaria Test Kits, Dental Implants, Dialysers, Electrosurgical Units, Endoscopes, Glucometer Test Strips, Guidewires, HIV Test Kits, Infusion Pumps, Intrauterine Devices, MRI Machines, Orthopaedic Implants, Patient Monitors, Surgical Staplers, Surgical Sutures, Ventilators, X-Ray Machines
Class D: Bone Cement, Coronary Stents, Heart Valves, Intraocular Lenses, Pacemakers, Spinal Implants
🔴 Class D Medical Devices (High risk)
Bone Cement — Class D
Bone cement is an implanted, load-bearing material that cures in the body and is central to joint fixation, which places it in the highest risk class (Class D) under the First Schedule.
- Key standards: ISO 5833, ISO 10993, ISO 5833., ISO 11135
- Device-specific documents: ISO 5833 test reports; Residual-monomer and working-time data; Biocompatibility (ISO 10993) reports
- Common CDSCO query: Mechanical (compressive/bending) or setting data short of ISO 5833.
Coronary Stents — Class D
A coronary stent is a long-term implantable device intended to be introduced into the central circulatory system and to remain there, which places it in the highest risk class under the First Schedule.
- Key standards: ISO 25539-2, ISO 5832, ISO 10993, ISO 11135
- Device-specific documents: Full design and mechanical test reports; Clinical investigation data; Biocompatibility (ISO 10993) reports for the stent, coating and any eluted drug
- Common CDSCO query: Clinical evidence that does not match the exact stent platform, size range or indication being registered.
Heart Valves — Class D
A prosthetic heart valve is a long-term, life-sustaining implant in the central circulatory system, which places it in the highest risk class (Class D) under the First Schedule.
- Key standards: ISO 5840-1, ISO 5840-2, ISO 5840-3, ISO 10993
- Device-specific documents: Hydrodynamic and durability test reports (ISO 5840); Clinical investigation data or accepted clinical-evaluation evidence for the specific valve; Biocompatibility (ISO 10993) and, for tissue valves, ISO 22442 animal-tissue safety/viral-inactivation data
- Common CDSCO query: Durability/fatigue data short of the labelled cycle life.
Intraocular Lenses — Class D
An intraocular lens is a long-term surgically invasive implant placed inside the eye, which places it in the highest risk class under the First Schedule.
- Key standards: ISO 11979-2, ISO 11979-3, ISO 11979-5, ISO 11979-7
- Device-specific documents: Optical and mechanical test reports (ISO 11979-2/-3); Clinical investigation data (ISO 11979-7) or accepted clinical-evaluation evidence for the specific lens; Biocompatibility (ISO 11979-5 / ISO 10993) for the lens material and any surface treatment
- Common CDSCO query: Clinical evidence that does not match the specific lens material or optical design.
Pacemakers — Class D
A pacemaker is a long-term active implant that sustains life, which places it in the highest risk class (Class D) under the First Schedule.
- Key standards: ISO 14708-1, ISO 14708-2, ISO 10993
- Device-specific documents: ISO 14708-2 test reports; Clinical investigation data or accepted clinical-evaluation evidence for the specific pacing system; Biocompatibility (ISO 10993) reports for the generator, header and leads
- Common CDSCO query: Clinical evidence not matching the specific pacing system or indication.
Spinal Implants — Class D
A spinal implant is a long-term, load-bearing surgical implant placed adjacent to the spinal cord, which places it in the highest risk class (Class D) under the First Schedule.
- Key standards: ISO 5832, ISO 14630, ISO 10993, ISO 11135
- Device-specific documents: Construct fatigue and static test reports (ASTM F1717 etc.) for the worst-case configuration; Cage compression/expulsion/subsidence data (ASTM F2077/F2267) for interbody devices; Clinical investigation / evaluation evidence for the specific system
- Common CDSCO query: Construct fatigue data not covering the worst-case configuration.
🟠 Class C Medical Devices (Moderate-to-high risk)
Anaesthesia Machines — Class C
An anaesthesia workstation is an active, life-supporting device controlling gas and agent delivery and ventilation, where failure is directly life-threatening, placing it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: ISO 80601-2-13, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8
- Device-specific documents: ISO 80601-2-13 performance report; IEC 60601-1 / -1-2 / -1-8 reports for electrical safety, EMC and alarms; Vaporiser and gas-monitoring evidence for integrated components
- Common CDSCO query: Gas/agent delivery accuracy or hypoxic-guard performance not evidenced to ISO 80601-2-13.
Blood Bags — Class C
A blood bag is a sterile blood-contacting device essential to transfusion safety, which places it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: ISO 3826, IS 14805, ISO 10993, ISO 11135
- Device-specific documents: ISO 3826 / IS 14805 test reports; Anticoagulant-solution specification and CoA against pharmacopoeial limits; Haemocompatibility and plasticiser (DEHP) data per ISO 10993
- Common CDSCO query: Closure/seal-integrity or centrifugation data missing.
Blood Gas Analysers — Class C
An IVD analyser whose results directly inform urgent critical-care decisions, where an incorrect result carries a significant patient risk, is classified at Class C under the IVD classification rules.
- Key standards: ISO 13485, ISO 23640, IEC 61010-2-101, IEC 60601-1-2
- Device-specific documents: Analytical performance reports; Cartridge/reagent and calibrator specifications and stability (ISO 23640); IVD equipment safety (IEC 61010-2-101) and EMC evidence
- Common CDSCO query: A reported analyte without accuracy/precision/linearity data.
Cardiac Catheters — Class C
A cardiac catheter is a surgically invasive device that connects to the central circulatory system, which places it in Class C (moderate-to-high risk) under the First Schedule — and electrophysiology/ablation and some structural-heart catheters sit at the upper bound.
- Key standards: ISO 10555-1, ISO 10555-4, ISO 11070, ISO 594
- Device-specific documents: ISO 10555 test reports; Sterilisation validation with the sterility assurance level (SAL 10⁻⁶); Haemocompatibility and biocompatibility (ISO 10993) for blood-contacting components and any hydrophilic coatin
- Common CDSCO query: Sterilisation validation missing or not matching the actual cycle.
COVID-19 Antigen Test Kits — Class C
An IVD that detects an infectious agent (SARS-CoV-2), where an incorrect result carries individual and public-health consequences, is classified at Class C under the IVD classification rules.
- Key standards: ISO 13485, ISO 23640, ISO 14971, ISO 18113
- Device-specific documents: Performance Evaluation Report from a designated laboratory; Analytical evidence; Stability (ISO 23640) data
- Common CDSCO query: Performance data not from a designated laboratory, or below the required sensitivity/specificity.
CT Scanners — Class C
A CT scanner is an active device delivering a significant ionising-radiation dose for diagnosis, whose malfunction can cause serious harm, placing it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: IEC 60601-2-44, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3
- Device-specific documents: IEC 60601-2-44 performance report; IEC 60601-1 / -1-2 / -1-3 reports for electrical safety, EMC and radiation protection; AERB type approval / no-objection for the CT X-ray source
- Common CDSCO query: IEC 60601-2-44 / -1-3 reports missing, out of date, or not for the exact model.
Defibrillators — Class C
An external defibrillator is an active life-supporting device delivering therapeutic energy in an emergency, where failure is directly life-threatening, placing it in Class C (moderate-to-high risk) under the First Schedule; implantable defibrillators are Class D.
- Key standards: IEC 60601-2-4, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8
- Device-specific documents: IEC 60601-2-4 performance report; IEC 60601-1 / -1-2 / -1-8 reports for electrical safety, EMC and alarms; Rhythm-analysis / shock-advisory algorithm validation for AEDs
- Common CDSCO query: Energy-accuracy or charge-time data not to IEC 60601-2-4.
Dengue & Malaria Test Kits — Class C
An IVD that detects an infectious agent (dengue virus or malaria parasite), where an incorrect result carries a significant risk to the patient, is classified at Class C under the IVD classification rules.
- Key standards: ISO 13485, ISO 23640, ISO 14971, ISO 18113
- Device-specific documents: Performance Evaluation Report from a designated laboratory; Analytical evidence; Stability (ISO 23640) data
- Common CDSCO query: Performance data not from a designated laboratory, or below required sensitivity/specificity.
Dental Implants — Class C
A dental implant is a long-term surgically invasive implant placed in bone, which places it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: ISO 14801, ISO 5832, ISO 10993, ISO 22794
- Device-specific documents: ISO 14801 dynamic-fatigue test report for the implant-abutment system; Material certificates to the relevant ISO 5832 grade; Biocompatibility (ISO 10993) reports for the implant, abutment and surface coating
- Common CDSCO query: ISO 14801 fatigue data missing or not covering the smallest/weakest configuration.
Dialysers — Class C
A dialyser is a blood-contacting device essential to a life-sustaining therapy, which places it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: ISO 8637-1, ISO 8637-2, ISO 10993, ISO 11135
- Device-specific documents: ISO 8637 performance reports; Haemocompatibility (ISO 10993) reports for the membrane and potting materials; Sterilisation validation with the sterility assurance level and residual limits
- Common CDSCO query: Clearance/ultrafiltration performance data missing or incomplete.
Electrosurgical Units — Class C
An electrosurgical unit is an active surgical device delivering high-frequency energy into the patient, where malfunction can cause burns or serious harm, placing it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: IEC 60601-2-2, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8
- Device-specific documents: IEC 60601-2-2 performance report; IEC 60601-1 / -1-2 / -1-8 reports for electrical safety, EMC and alarms; Accessory evidence for electrodes, pencils and return pads
- Common CDSCO query: HF-output accuracy or return-electrode monitoring not evidenced to IEC 60601-2-2.
Endoscopes — Class C
An endoscope is a surgically invasive device that contacts internal tissue, typically reused, where inadequate reprocessing carries a cross-infection risk, placing it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: ISO 8600, ISO 10993, ISO 17664, IEC 60601-1
- Device-specific documents: ISO 8600 performance reports; Reprocessing validation (ISO 17664); Biocompatibility (ISO 10993) for patient-contacting parts
- Common CDSCO query: Reprocessing instructions not validated (ISO 17664) or not hospital-executable.
Glucometer Test Strips — Class C
An IVD for self-testing that measures blood glucose to manage a chronic condition, where an incorrect result poses a significant risk to the patient, is classified at Class C under the IVD classification rules.
- Key standards: ISO 15197, ISO 13485, ISO 23640, ISO 14971
- Device-specific documents: ISO 15197 system-accuracy and user-performance study for the meter-and-strip system; Performance Evaluation Report; Stability (ISO 23640) data
- Common CDSCO query: ISO 15197 accuracy study that does not meet the current edition's limits, or is not a user-performance study.
Guidewires — Class C
A guidewire is a surgically invasive device that connects to the central circulatory system for short-term use, placing it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: ISO 11070, ISO 10993, ISO 11135
- Device-specific documents: ISO 11070 test reports; Sterilisation validation with the sterility assurance level (SAL 10⁻⁶); Haemocompatibility (ISO 10993) for the wire and coating
- Common CDSCO query: Sterilisation validation missing or not matching the actual cycle.
HIV Test Kits — Class C
An IVD that detects an infectious agent (HIV) where an incorrect result poses a significant risk to the individual and to public/transfusion safety is classified at Class C under the IVD classification rules.
- Key standards: ISO 13485, ISO 23640, ISO 14971, ISO 18113
- Device-specific documents: Performance Evaluation Report from a CDSCO-designated referral laboratory; Reference-panel results including seroconversion panels where applicable; Stability (ISO 23640) data
- Common CDSCO query: Performance data not from a CDSCO-designated referral laboratory, or below the required sensitivity/specificity.
Infusion Pumps — Class C
An infusion pump is an active device that controls the delivery of fluids or medication, where a delivery error can cause serious harm, placing it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: IEC 60601-2-24, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8
- Device-specific documents: IEC 60601-2-24 performance report; IEC 60601-1 / -1-2 reports for electrical safety and EMC, and IEC 60601-1-8 alarm evidence; Software / firmware documentation
- Common CDSCO query: Flow-accuracy or occlusion-alarm data not to IEC 60601-2-24, or not for the exact model and sets.
Intrauterine Devices — Class C
An IUD is a long-term device introduced into the uterus, which places it in Class C (moderate-to-high risk) under the First Schedule; a hormonal IUD is a drug-device combination because its contraceptive action also depends on a released medicinal substance.
- Key standards: ISO 10993, ISO 11135
- Device-specific documents: Biocompatibility (ISO 10993) reports for long-term uterine contact; Copper-release / dissolution data demonstrating performance over the labelled lifetime (copper IUDs); Mechanical and insertion-system test reports
- Common CDSCO query: Copper-release/dissolution data missing or not supporting the claimed lifetime.
MRI Machines — Class C
An MRI machine is an active diagnostic device with strong static and gradient magnetic fields and RF energy, whose malfunction or mismanagement can cause serious harm, placing it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: IEC 60601-2-33, IEC 60601-1, IEC 60601-1-2, ISO 10993
- Device-specific documents: IEC 60601-2-33 performance report; IEC 60601-1 / -1-2 reports for electrical safety and EMC; Quench and cryogen-management evidence for superconducting systems
- Common CDSCO query: IEC 60601-2-33 report missing, incomplete or not for the exact model/field strength.
Orthopaedic Implants — Class C
A bone plate or screw is a long-term surgically invasive implant that bears load inside the body, which places it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: ISO 5832, ISO 14630, ISO 5838, ISO 9268
- Device-specific documents: Material certificates to the relevant ISO 5832 grade for every implant; Mechanical test reports; Biocompatibility (ISO 10993) reports for the implant material and any coating
- Common CDSCO query: Material certificates that do not tie to the specific ISO 5832 grade claimed.
Patient Monitors — Class C
A patient monitor is an active device whose measurements and alarms directly inform urgent clinical decisions, where failure can be life-threatening, placing it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: IEC 60601-2-27, IEC 60601-2-49, ISO 80601-2-61, IEC 80601-2-30
- Device-specific documents: Parameter-specific performance reports; IEC 60601-1 / -1-2 / -1-8 reports for electrical safety, EMC and alarms; Software / firmware documentation
- Common CDSCO query: A measured parameter claimed without its particular-standard performance report.
Surgical Staplers — Class C
A surgical stapler performs a critical tissue-joining function and leaves implanted staples in the body, which places it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: ISO 5832, ISO 10993, ISO 11135, IEC 60601-1
- Device-specific documents: Mechanical test reports; Implant-material certificates and biocompatibility (ISO 10993) for the staples; Sterilisation validation with the sterility assurance level
- Common CDSCO query: Staple-line integrity or staple-formation data missing.
Surgical Sutures — Class C
A surgical suture is a surgically invasive implant used to close tissue, where absorbable and long-dwell sutures carry implant-level risk, which places it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: ISO 10993, ISO 11135
- Device-specific documents: Pharmacopoeial test reports; Absorption-profile data for absorbable sutures (tensile-retention and mass-loss over time); Biocompatibility (ISO 10993) reports for the material, dye and coating
- Common CDSCO query: Tensile (knot-pull) or diameter results outside the pharmacopoeial limits for a gauge.
Ventilators — Class C
A ventilator is an active, life-supporting device whose failure can directly cause death or serious deterioration, placing it in Class C (moderate-to-high risk) — and some life-support critical-care ventilators are treated at the higher end of the scale.
- Key standards: ISO 80601-2-12, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8
- Device-specific documents: IEC 60601-1 electrical-safety test report from an accredited laboratory; IEC 60601-1-2 EMC report and the IEC 60601-1-8 alarm-system evidence; ISO 80601-2-12 performance report covering ventilation modes, accuracy and alarms
- Common CDSCO query: IEC 60601-1 / -1-2 reports that are out of date, incomplete, or not for the exact model.
X-Ray Machines — Class C
An X-ray machine is an active device that emits ionising radiation for diagnosis, whose malfunction can cause significant harm, placing it in Class C (moderate-to-high risk) under the First Schedule.
- Key standards: IEC 60601-2-54, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3
- Device-specific documents: IEC 60601-2-54 performance report; IEC 60601-1 / -1-2 / -1-3 reports for electrical safety, EMC and radiation protection; AERB type approval / no-objection for the X-ray source and tube
- Common CDSCO query: IEC 60601-2-54 / -1-3 reports missing, out of date, or not for the exact model.
🟡 Class B Medical Devices (Low-to-moderate risk)
Autoclaves & Sterilizers — Class B
A sterilizer is a device whose failure leaves instruments non-sterile and creates an infection risk, but which does not itself contact the patient, placing it in Class B (low-to-moderate risk) under the First Schedule.
- Key standards: ISO 17665, IEC 61010-2-040, IEC 60601-1, IEC 60601-1-2
- Device-specific documents: Cycle-validation reports (ISO 17665); EN 285 / EN 13060 type-test evidence as applicable to size; Electrical and steam-safety reports (IEC 61010-2-040)
- Common CDSCO query: Cycle-validation (ISO 17665) evidence missing or incomplete.
Blood Pressure Monitors — Class B
A blood pressure monitor has a measuring function whose reading informs diagnosis and treatment, and CDSCO classifies devices with a measuring function at Class B under the First Schedule.
- Key standards: IEC 80601-2-30, ISO 81060-2, ISO 81060-1, IEC 60601-1
- Device-specific documents: ISO 81060-2 clinical-accuracy validation; IEC 80601-2-30 performance report and IEC 60601-1 / -1-2 electrical safety/EMC; Cuff-range and calibration data across the arm/wrist sizes offered
- Common CDSCO query: No ISO 81060-2 clinical-accuracy validation for an automated monitor.
Condoms — Class B
A condom is a barrier device used briefly that prevents pregnancy and infection, where failure carries a real health consequence, placing it in Class B under the First Schedule.
- Key standards: ISO 4074, IS 3701, ISO 23409, ISO 10993
- Device-specific documents: ISO 4074 / IS 3701 test reports; Biocompatibility and sensitisation data for the latex and lubricant; Lubricant / additive composition and safety for lubricated or flavoured variants
- Common CDSCO query: Burst-volume/pressure or freedom-from-holes results outside ISO 4074 / IS 3701 limits.
Contact Lenses — Class B
A contact lens is a device in prolonged contact with the eye surface, where material and hygiene issues carry an ocular-health risk, placing it in Class B (low-to-moderate risk) under the First Schedule.
- Key standards: ISO 18369, ISO 18369-4, ISO 9394, ISO 10993
- Device-specific documents: ISO 18369 test reports; Biocompatibility (ISO 9394 / ISO 10993) for the lens material and, for coloured lenses, the pigments; Colour-additive safety and leaching data for coloured/cosmetic lenses
- Common CDSCO query: Oxygen-permeability (Dk/t) or material data missing.
Digital Thermometers — Class B
A clinical thermometer has a measuring function whose reading informs diagnosis, and CDSCO classifies devices with a measuring function at Class B under the First Schedule.
- Key standards: ISO 80601-2-56, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11
- Device-specific documents: ISO 80601-2-56 performance report; IEC 60601-1 / -1-2 reports for electrical safety and EMC; Calibration and accuracy data across the measurement range
- Common CDSCO query: Accuracy/performance data not to ISO 80601-2-56, or across only part of the range.
Disposable Syringes — Class B
A syringe is a surgically invasive device for transient use that delivers into the body, which places it in Class B (low-to-moderate risk) under the First Schedule.
- Key standards: IS 10258, ISO 7886-1, IS/ISO 7864, ISO 11135
- Device-specific documents: IS 10258 / ISO 7886 test reports; Sterilisation validation; Biocompatibility reports for the barrel, plunger and lubricant
- Common CDSCO query: Sterilisation validation missing, out of date, or not matching the actual cycle used.
ECG Machines — Class B
An ECG machine is an active diagnostic device connected to the patient whose output informs clinical decisions, placing it in Class B under the First Schedule.
- Key standards: IEC 60601-2-25, IEC 60601-2-27, IEC 60601-1, IEC 60601-1-2
- Device-specific documents: IEC 60601-2-25 performance report; IEC 60601-1 / -1-2 reports for electrical safety and EMC; Software / firmware documentation where the device interprets or measures the ECG
- Common CDSCO query: IEC 60601-2-25 performance data missing or not for the exact model.
Enteral Feeding Tubes — Class B
An enteral feeding tube is a surgically invasive device introduced through a body orifice (or stoma) that connects to the GI tract, placing it in Class B (low-to-moderate risk) under the First Schedule; long-term gastrostomy devices may be classified higher.
- Key standards: ISO 20695, ISO 80369-3, ISO 10993, ISO 11135
- Device-specific documents: ISO 20695 test reports; ISO 80369-3 (ENFit) connector conformity to prevent misconnection; Sterilisation validation with the sterility assurance level
- Common CDSCO query: Wrong connector — a luer instead of the ENFit (ISO 80369-3) enteral connector.
Hearing Aids — Class B
A hearing aid is a device that compensates for a disability and whose sound output affects the user, placing it in Class B (low-to-moderate risk) under the First Schedule.
- Key standards: IEC 60118, IEC 60118-0, IEC 60118-7, IEC 60601-1
- Device-specific documents: IEC 60118 performance reports; Electrical-safety / EMC and battery-charging evidence where relevant; Biocompatibility for ear-canal-contacting parts
- Common CDSCO query: IEC 60118 performance data missing or not for the exact model.
Hospital Beds — Class B
A medical hospital bed with powered adjustment and side rails carries entrapment, pinch and electrical risks, placing it in Class B (low-to-moderate risk) under the First Schedule; a basic non-adjustable bed may be Class A.
- Key standards: IEC 60601-2-52, IEC 60601-1, IEC 60601-1-2, ISO 10993
- Device-specific documents: IEC 60601-2-52 test reports; IEC 60601-1 / -1-2 reports for electrical safety and EMC of the actuators; Mechanical/fatigue test data for the frame and lift mechanism
- Common CDSCO query: Entrapment-zone (side-rail) dimensions not evidenced to IEC 60601-2-52.
Insulin Pens — Class B
An insulin pen is a device that delivers a measured dose of medication, whose dose accuracy affects patient safety, placing it in Class B (low-to-moderate risk) under the First Schedule; a prefilled pen with insulin is a drug-device combination.
- Key standards: ISO 11608-1, ISO 11608-2, ISO 11608-3, ISO 7864
- Device-specific documents: ISO 11608 dose-accuracy test reports across the dose range and conditions; Pen-needle sterility and sharpness/penetration data (ISO 7864); Mechanical reliability / dose-setting evidence over the pen's life
- Common CDSCO query: Dose-accuracy data not to ISO 11608 across the full range.
IV Cannula — Class B
An IV cannula is a surgically invasive device that connects to the circulatory system for short-term use, placing it in Class B (low-to-moderate risk) under the First Schedule.
- Key standards: ISO 10555-1, ISO 10555-5, ISO 594, ISO 80369-7
- Device-specific documents: ISO 10555 test reports; Sterilisation validation with the sterility assurance level (SAL 10⁻⁶); Biocompatibility reports for the catheter, hub and any port
- Common CDSCO query: Sterilisation validation missing or not matching the actual cycle.
Nebulizers — Class B
A nebulizer is an active device that delivers medication to the respiratory tract, whose output affects dosing, placing it in Class B under the First Schedule.
- Key standards: ISO 27427, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11
- Device-specific documents: ISO 27427 performance report; IEC 60601-1 / -1-2 reports for electrical safety and EMC; Biocompatibility for the mask, mouthpiece and medication chamber
- Common CDSCO query: Particle-size / aerosol-output performance not evidenced to ISO 27427.
Oxygen Concentrators — Class B
An oxygen concentrator is an active therapeutic device that supplies a medical gas, whose performance affects the patient, placing it in Class B under the First Schedule.
- Key standards: ISO 80601-2-69, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11
- Device-specific documents: ISO 80601-2-69 performance report; IEC 60601-1 / -1-2 reports for electrical safety and EMC; Alarm evidence for low-oxygen-concentration and power/flow failure
- Common CDSCO query: Oxygen-concentration/purity performance not evidenced across the full flow range.
Pregnancy Test Kits — Class B
An hCG pregnancy test is a self-testing IVD that gives a preliminary qualitative result normally confirmed by other means, which places it at Class B (moderate risk) under the IVD classification rules.
- Key standards: ISO 13485, ISO 23640, ISO 14971, ISO 18113
- Device-specific documents: Performance Evaluation Report; Stability (ISO 23640) data; Cut-off / detection-limit justification for the stated hCG level
- Common CDSCO query: Performance data that does not support the claimed sensitivity/specificity or detection limit.
Surgical Gloves — Class B
A surgical glove is a sterile, single-use device that contacts breached skin and internal tissue during surgery, placing it in Class B under the First Schedule.
- Key standards: IS 4148, ISO 10282, IS 15354, ISO 11193
- Device-specific documents: IS 4148 / ISO 10282 test reports; Sterilisation validation for sterile gloves, with the sterility assurance level; Biocompatibility and protein-content data for latex gloves (to manage sensitisation risk)
- Common CDSCO query: AQL / freedom-from-holes results missing or above the acceptable limit.
Ultrasound Machines — Class B
An ultrasound machine is an active diagnostic device using non-ionising acoustic energy, whose output affects diagnosis, placing it in Class B (low-to-moderate risk) under the First Schedule.
- Key standards: IEC 60601-2-37, IEC 60601-1, IEC 60601-1-2, ISO 10993
- Device-specific documents: IEC 60601-2-37 performance report; IEC 60601-1 / -1-2 reports for electrical safety and EMC; Transducer specifications and biocompatibility for each probe
- Common CDSCO query: Acoustic-output (MI/TI) data not demonstrated to IEC 60601-2-37.
Urinary Catheters — Class B
A urinary catheter is a surgically invasive device introduced through a body orifice for short-to-medium-term use, placing it in Class B under the First Schedule.
- Key standards: ISO 20696, ISO 20697, ISO 10993, ISO 11135
- Device-specific documents: ISO 20696 test reports; Sterilisation validation with the sterility assurance level (SAL 10⁻⁶); Biocompatibility reports for the catheter and coating (e.g. silicone, hydrogel)
- Common CDSCO query: Balloon-integrity or inflation data missing for Foley catheters.
Wound Dressings — Class B
An advanced wound dressing is a device that contacts breached skin and manages the wound environment for a limited period, which places it in Class B (low-to-moderate risk) under the First Schedule; simple non-medicated dressings can be Class A, and medicated dressings sit higher.
- Key standards: ISO 10993, ISO 11135
- Device-specific documents: EN 13726 test reports; Biocompatibility (ISO 10993) reports for prolonged wound contact; Sterilisation validation where supplied sterile, with the sterility assurance level
- Common CDSCO query: Fluid-handling/absorbency data missing or not matching the dressing type.
🟢 Class A Medical Devices (Low risk)
Examination Gloves — Class A
An examination glove is a low-risk device providing a protective barrier for general, non-surgical use, placing it in Class A under the First Schedule.
- Key standards: IS 15354, ISO 11193-1, ISO 11193-2, ISO 10993
- Device-specific documents: IS 15354 / ISO 11193 test reports; Biocompatibility and protein-content data for latex gloves; Powder status declaration
- Common CDSCO query: AQL / freedom-from-holes results missing or above the acceptable limit.
Surgical Masks — Class A
A surgical mask is a low-risk device that provides a protective barrier and is used briefly and externally, placing it in Class A under the First Schedule.
- Key standards: IS 16289, ISO 10993, ISO 11135
- Device-specific documents: BFE test report; Differential-pressure (breathability) report and, for Type IIR, splash-resistance data; Particle filtration and microbial-cleanliness data as required by the standard
- Common CDSCO query: BFE or differential-pressure results missing or below the declared type.
❓ Frequently Asked Questions
How does CDSCO classify a medical device?
Under the First Schedule of the Medical Devices Rules, 2017, every device is placed in Class A (low risk), B (low-moderate), C (moderate-high) or D (high risk) based on its intended use, duration of contact, invasiveness and whether it is active.
Which form do I file to import a medical device into India?
An Indian Authorised Agent holding a manufacturing or wholesale licence files Form MD-14 with the Central Licensing Authority; the import licence is issued in Form MD-15.
Who licenses domestic manufacture?
Class A and B devices are licensed by the State Licensing Authority (Form MD-3 application, MD-5 licence). Class C and D devices are licensed by the Central Licensing Authority (Form MD-7 application, MD-9 licence).
Does a CDSCO medical device licence expire?
No. Licences under the Medical Devices Rules, 2017 are perpetual, provided the retention fee is paid every five years.
How long does CDSCO device registration take?
A complete, well-prepared manufacturing file typically clears in 3–6 months and an import licence in 6–9 months. Every deficiency letter restarts the clock, so the quality of the first submission matters most.