CDSCO · Class C IVD

CDSCO Licence for Blood Gas Analysers in India

Class C IVD licensing for blood gas / electrolyte analysers — performance evaluation, documents, fees, timeline and routes.

Published: September 2026 Reading Time: 13 min

A blood gas analyser measures pH, pO₂, pCO₂ and often electrolytes and metabolites from a blood sample, driving urgent decisions in the ICU, OT and emergency room. Because it is an in-vitro diagnostic whose results guide critical care, CDSCO regulates the blood gas analyser as a Class C IVD with a performance-evaluation report behind the licence. This guide covers the full CDSCO pathway for blood gas analysers under the Medical Devices Rules, 2017 — classification, licence route, the performance evidence, the documents, the fees and the timeline.

📌 Under the Medical Devices Rules, 2017, Blood Gas Analysers are a notified medical device and cannot be manufactured, imported or sold in India without a CDSCO licence. As a Class C device (moderate-to-high risk), a domestic manufacturer applies in Form MD-7 to the Central Licensing Authority (CDSCO headquarters) and receives the licence in Form MD-9; an importer applies in Form MD-14 through an Indian Authorised Agent and receives an import licence in Form MD-15 from the Central Licensing Authority. The licence must be in hand before the product is made or the first consignment lands.

📋 Is a CDSCO Licence Mandatory for Blood Gas Analysers?

Yes. A blood gas analyser is a notified in-vitro diagnostic and cannot be manufactured, imported or sold in India without a CDSCO licence. Because its measurements are used in critical-care decisions, it is a Class C IVD: a domestic manufacturer applies to the Central Licensing Authority in Form MD-7 for a Form MD-9 licence, and an importer applies in Form MD-14 for a Form MD-15 import licence. Benchtop and point-of-care analysers, and their cartridges/reagents, fall within the framework.

🧭 Risk Classification of Blood Gas Analysers

Every medical device in India is placed in one of four risk classes under the First Schedule of the Medical Devices Rules, 2017. The class is the single most important decision in the whole application — it fixes which authority you file with, which form you use, the fee, the audit route and the timeline.

Class ALow risk
Class BLow-moderate
Class CModerate-high
Class DHigh risk

Blood Gas Analysers are classified as Class C. An IVD analyser whose results directly inform urgent critical-care decisions, where an incorrect result carries a significant patient risk, is classified at Class C under the IVD classification rules. Classification follows the CDSCO classification list for the relevant medical-device category read with the First Schedule rules; where a device could fall in two classes, the higher class applies, and an accessory is classified in its own right.

🪪 Which CDSCO Licence You Need for Blood Gas Analysers

There are two distinct routes, and which one applies depends on whether you make the device in India or bring it in from overseas:

📘 Applicable Standards for Blood Gas Analysers

Blood gas analysers are evaluated principally on performance evidence, supported by IVD quality and electro-medical standards:

🗂️ The Core Technical Dossier — Required for Every Application

Whether you file for a manufacturing licence or an import licence, CDSCO expects the same backbone of technical evidence built around the device. Getting this base right is what decides whether the file clears in one review cycle or comes back as a deficiency letter; the device-specific documents in the next section sit on top of it.

➕ Device-Specific Documents for Blood Gas Analysers

On top of the core dossier, a blood-gas-analyser IVD file should carry:

For an import licence (Form MD-14) there are three further essentials on top of the dossier above:

🏭 QMS & Essential Principles — What CDSCO Checks

The Medical Devices Rules, 2017 are built on two pillars, and both are assessed for blood gas analysers:

🏷️ Blood Gas Analysers Label & IFU Requirements

IVD labelling (ISO 18113) is checked for analytes, lot and calibration information:

⚖️ Analyser and Consumables: One System

A blood gas analyser is an instrument plus its cartridges, reagents and calibrators, and the analytical performance evidence ties them together — the instrument alone is not the whole product. Each analyte reported (pH, gases, electrolytes, metabolites) needs its own performance data, and the consumables need stability and lot data. Register the analyser with its defined consumables and evidence each reported parameter — a parameter claimed without performance data is the classic query.

🛠️ Step-by-Step Registration Process

  1. Confirm the classificationFix the risk class against the CDSCO classification list and the First Schedule — this decides the form, the authority, the fee and the audit route before anything else is drafted.
  2. Choose the routeManufacturing (Form MD-3 / MD-7) or import (Form MD-14). Importers appoint an Indian Authorised Agent holding a valid manufacturing or wholesale licence.
  3. Build the Device Master File and Plant Master FileAssemble device description, materials, design and process data, ISO 14971 risk file, sterilisation and stability data, and the site file with the QMS.
  4. Complete the Essential Principles checklistMap the device against each applicable essential principle and cite the BIS/IS or ISO/IEC standard or test report relied on for each.
  5. Draw the test and performance reportsTest to the applicable standards at an accredited laboratory; sterile and measuring devices carry additional validation.
  6. Register and file on the CDSCO Online (SUGAM) portalCreate the account, pay the government fee online and upload the full dossier under the correct form.
  7. Site audit / inspectionNotified-Body audit for Class A/B manufacture; CDSCO joint-team inspection for Class C/D and, where ordered, for overseas sites on import.
  8. Respond to queries and receive the licenceAnswer the deficiency letter within the stipulated period; on approval the licence issues in Form MD-5 / MD-9 (manufacture) or MD-15 (import).

💰 Government Fees

Fees are set in the Second Schedule of the Medical Devices Rules and paid online through the CDSCO portal. They differ sharply by route and class:

Manufacture — Class C (Form MD-9)

Rs 50,000 per site plus Rs 1,000 for each distinct device, to the Central Licensing Authority (CDSCO headquarters).

Import (Form MD-15)

Charged per site and per device and paid in USD; for a Class C device it is among the higher import slabs, so budget by the number of sites and distinct devices.

Test licence (Form MD-13)

A modest fee for importing or making a small quantity for testing, evaluation, demonstration or clinical investigation.

Retention

Licences do not expire, but a retention fee equal to the licence fee is payable every five years to keep the licence alive.

Import fees under Form MD-15 run from roughly USD 1,000 per site plus USD 50 per device for lower-risk devices to about USD 3,000 per site plus USD 1,500 per device for Class C and D. Always confirm the current figure on the CDSCO portal before you file, as the schedule is revised from time to time.

⏱️ Timeline & Validity at a Glance

6-9 monthsManufacture (Form MD-9)
6-9 monthsImport (Form MD-15)
PerpetualLicence Validity
5 yearsRetention Fee Cycle

The single biggest lever on timeline is the quality of the first submission: every deficiency letter effectively restarts the clock, so a dossier that pre-empts the common objections is worth far more than one filed quickly and corrected later. Unlike the old registration certificates, an MDR 2017 licence does not carry an expiry date — it stays valid as long as the retention fee is paid every five years and the conditions of licence are met.

⚠️ Where Blood Gas Analysers Applications Get Queried

Common query reasons for blood-gas-analyser IVD applications are:

🔗 Related Approvals Alongside CDSCO

A CDSCO device licence rarely travels alone. Depending on the device and the business, plan for these adjacent approvals in parallel, because any one of them can hold a consignment at the port or stop a sale even when the CDSCO licence is in order:

✅ Post-Approval Obligations

The licence is the start of an ongoing compliance duty, not the end of it. After grant, the licensee must:

Confirm the current position before you file: This guide reflects the Medical Devices Rules, 2017 and CDSCO guidance as published and is current as at 2026. Forms, fees, classification and standards are revised from time to time — verify the applicable class, standard and fee on the CDSCO portal, or with us, before drawing test reports or filing.

❓ Frequently Asked Questions

What CDSCO class is a blood gas analyser?

A blood gas analyser is a Class C IVD, so a domestic manufacturer files Form MD-7 for an MD-9 licence with the Central Licensing Authority, and an importer files Form MD-14 for an MD-15 licence.

What performance evidence does CDSCO expect?

Accuracy, precision, linearity and reportable range for each analyte (pH, pO₂, pCO₂, electrolytes), with calibration traceability and consumable stability.

Are the cartridges and reagents part of the registration?

Yes — the analyser is registered as a system with its defined cartridges, reagents and calibrators, and their stability and lot data are part of the evidence.

Do IVDs use different forms?

No — IVDs use the same MDR 2017 forms (MD-9 for Class C manufacture, MD-15 for import); the distinguishing requirement is the performance-evaluation evidence.

How long does registration take?

Budget roughly six to nine months for a complete Class C IVD manufacturing (MD-9) or import (MD-15) file.

Does the licence expire?

No — the MDR 2017 licence is perpetual, subject to the five-yearly retention fee and continued compliance.

Blood Gas Analysers — CDSCO Licensing Handled End to End

Global Approbation manages the complete CDSCO medical-device process for blood gas analysers — classification, standard mapping, Device and Plant Master File preparation, ISO 13485 and Essential Principles readiness, SUGAM filing, audit support and query response — so your licence issues without avoidable delay.